Validation of a Commercial Dry-Form Broth Microdilution Device (Sensititre) for Testing Tedizolid, a New Oxazolidinone
Authors:
Ronald N. Jones, N. M. Holliday, Paul R. Rhomberg
Abstract:ABSTRACTTedizolid, a novel oxazolidinone antibacterial with potent activity against a wide range of Gram-positive pathogens, was recently approved by regulatory authorities for the treatment of acute bacterial skin and skin structure infections. A commercial broth microdilution device (Sensititre; Thermo Fisher Scientific) was validated using 285 selected Gram-positive isolates, and the device was documented to have 100.0% essential and categorical agreement with reference MIC results and excellent MIC endpoint (...) ABSTRACTTedizolid, a novel oxazolidinone antibacterial with potent activity against a wide range of Gram-positive pathogens, was recently approved by regulatory authorities for the treatment of acute bacterial skin and skin structure infections. A commercial broth microdilution device (Sensititre; Thermo Fisher Scientific) was validated using 285 selected Gram-positive isolates, and the device was documented to have 100.0% essential and categorical agreement with reference MIC results and excellent MIC endpoint reproducibility.(Read More)